Pre-flight Calibration and validation test of CMM-1 Fluxgate Magnetometer MAG-O
- 1University of Science and Technology of China, Institute of earth and space science, Department of Geophysics and Planetary Science, China (aiwenzi@mail.ustc.edu.cn)
- 2Space Research Institute, Austrian Academy of Sciences, Graz, Austria
As one of the objectives for China’s First Mars Exploration Mission (CMM-1), the magnetic field in Mars’ near space will be measured by the Magnetometer onboard of the Orbiter (MAG-O), which consists of two fluxgate sensors and one electronic box. We conducted pre-flight calibration test to determine the temperature dependent offset and the correction matrix, which composed of sensitivities and non-orthogonality parameters. Two sets of ground validation experiments were also executed under limited condition to confirm the consistency and mutual interference of the two sensors, respectively. The results show that MAG-O meets the requirements of CMM-1 mission. This paper will introduce the procedure and results of pre-flight calibration and validation test.
How to cite: Hu, X., Liu, K., Li, X., Pan, Z., Li, Y., Hao, X., and Zhang, T.: Pre-flight Calibration and validation test of CMM-1 Fluxgate Magnetometer MAG-O, EGU General Assembly 2020, Online, 4–8 May 2020, EGU2020-2090, https://doi.org/10.5194/egusphere-egu2020-2090, 2020